Avicenna J Dent Res. 15(2):47-52.
doi: 10.34172/ajdr.2023.1624
Original Article
Effect of Diamond-like Carbon Coating Applied by the Physical Vapor Deposition Technique on Wear of Diamond and Tungsten Carbide Dental Burs
Bahareh Asgartooran 1
, Zahra Khamverdi 1
, Faezeh Soury 2, Shiva Kavousinejad 1, * 
Author information:
1Department of Restorative Dentistry, Dental School, Hamadan University of Medical Sciences, Hamadan, Iran
2Dental School, Hamadan University of Medical Sciences, Hamadan, Iran
Abstract
Background: The methods of increasing the longevity of dental burs by improving the mechanical properties of these surfaces, can increase their longevity. This study assessed the effect of diamond-like carbon (DLC) coating applied by the physical vapor deposition (PVD) technique on wear of diamond and tungsten carbide (TC) burs.
Methods: In this in vitro study, 30 diamond and 30 TC burs were evaluated in four groups, including TC burs without coating (control), TC burs with a 3.5-µm DLC coating applied by the PVD technique, diamond burs without coating (control), and diamond burs with a 3.5-µm DLC coating applied by the PVD technique. The burs were weighed by a digital scale, underwent the pin-on-disc wear test, and were weighed again. The weight loss indicated the degree of wear in each group. For qualitative assessments, the surface of the burs was inspected under a stereomicroscope at×4 and×10 magnifications before wear, halfway through the test, and after the test. Finally, the data were analyzed by two-way ANOVA (α=0.05).
Results: The effect of DLC coating was significant on the wear of burs (P=0.032), but the effect of the type of bur and their interaction effect on wear were not significant (P=0.151). A significant difference existed in wear among the four groups (P < 0.001), and the wear of coated burs was significantly lower than that of non-coated burs (P=0.012). Stereomicroscopic assessments revealed some residual diamond particles, the impression of dislodged particles and the path of wear on the surface of diamond burs, and the path of wear on the surface of TC burs.
Conclusions: Overall, the DLC coating of diamond and TC dental burs by the PVD technique could increase their wear resistance irrespective of the bur type.
Keywords: Diamond, Tungsten carbide, Dental burs, Physical vapor deposition
Copyright and License Information
© 2023 The Author(s); Published by Hamadan University of Medical Sciences.
This is an open-access article distributed under the terms of the Creative Commons Attribution License (
http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium provided the original work is properly cited.
Please cite this article as follows: Asgartooran B, Khamverdi Z, Soury F, Kavousinejad S.Effect of diamond-like carbon coating applied by the physical vapor deposition technique on wear of diamond and tungsten carbide dental burs. Avicenna J Dent Res. 2023; 15(2):47-52. doi:10.34172/ajdr.2023.1624
Background
Dental burs are routinely used for sectioning and removal of hard tissues such as the tooth structure and bone. The three main components of dental burs include the body, shaft, and tip (the cutting part) (1,2). Burs are designed in different shapes and sizes and can operate at a speed of 500 000 rpm in dental hand-pieces. Dental burs are made of steel, steel, tungsten carbide (TC), pure TC, or stainless steel, and diamond particles. Due to the high hardness of TC, TC burs are highly resistant to wear. Thus, blades and the body part are often made of TC and steel, respectively. Further, the addition of 6% cobalt to TC increases its resistance. Diamond burs are composed of diamond particles attached to a metal base. Therefore, they have drawbacks such as heterogeneous size and shape of particles, difficult sterilization, and short longevity and clinical service (3-5). Dental burs have been employed in dentistry for over 300 years (5), and many attempts have been made to improve their physical properties (6).
Cutting efficiency is an important physical property for burs. However, due to the friction of the cutting edge, burs undergo wear and degradation (3). Furthermore, in the process of repeated sterilization and disinfection cycles, the superficial cutting layers corrode and become dull (7). The use of dull and corroded burs increases the pulpal temperature in dental procedures (8). Pulpal temperature rise has always been a topic of debate in dental research (9,10). In case of temperature rise > °5, the risk of irreversible pulpitis increases by 10%. This rate increases to over 60% in case of a temperature rise over 11°C (11).
TC instruments are among the most commonly used cutting tools. To improve their properties and increase their longevity, TC burs are often coated with titanium nitride, titanium carbon nitride, and aluminum oxide. These coatings can be employed in one single or multiple layers. Coatings increase the wear resistance, hardness, and heat resistance of instruments, and improve the quality of tooth preparation due to lower friction and increased dimensional stability of the coated instrument. For example, aluminum oxide coating increases the wear resistance and chemical stability of instruments at high temperatures (12). Diamond coatings are also routinely applied on cutting instruments. Diamond coatings improve the function and properties of cutting instruments (13). On the other hand, although the diamond coating is a well-accepted technique, optimal coating for dental burs has not been well investigated yet (14).
Chemical vapor deposition (CVD) and physical vapor deposition (PVD) are the frequently used coating techniques, which are performed at low temperatures (15). At present, plasma-assisted CVD is considered to apply hard corrosion-resistant and wear-resistant coatings such as diamond-like carbon (DLC) on different tools (16). However, the high vapor pressure required for the CVD technique and residual graphite are the main drawbacks of this technique. Currently, the production of industrial diamonds is routinely performed, and many laboratories produce high-purity diamonds (17). The PVD technique applies titanium aluminum nitride coatings with excellent adhesion properties, along with high wear resistance. Thus, it is commonly applied to create a thermal barrier and increase wear resistance (18).
However, low ductility and poor adhesion to the underlying structure due to high residual stresses in the coating are the main limitations against the widespread use of DLC coating. Evidence suggests that the addition of elements such as silicium and chromium can obviate these limitations to some extent. Additionally, a suitable intermediate layer can be applied between the DLC coating and the underlying structure (19). At present, DLC coatings are increasingly used in medicine and dentistry. They are applied over biocompatible dental implants to ensure their optimal wear and corrosion resistance and hardness, as well as the low friction coefficient (20,21). DLC coatings are also applied over the stainless steel tip of ultrasonic instruments for bone cutting to minimize the temperature rise (22). DLC coatings are applied over nickel-titanium orthodontic wires to improve their wear and corrosion resistance. Moreover, these coatings prevent the release of nickel ions from the wires (23). They are highly efficient for cutting instruments due to their high hardness and low friction coefficient. This process can improve their wear resistance and longevity, prevent their corrosion and rusting, and enhance their function as such. It has been reported that the coating of instruments improves their quality of function by 300% (24).
Considering all the above-mentioned explanations, this study aimed to assess the effect of DLC coating applied by the PVD technique on the wear of diamond and TC dental burs.
Materials and Methods
This in vitro, experimental study was conducted on 30 diamond and 30 TC burs. The sample size was calculated to be 15 in each group assuming alpha = 0.05 and a study power of 80% (14).
The inclusion criteria were sound burs with no defects or cracks and the correct application of coating. The burs were evaluated in four groups (n = 15), including TC burs without coating (control), TC burs with a 3.5-µm DLC coating applied by the PVD technique, diamond burs without coating (control), and diamond burs with a 3.5-µm DLC coating applied by the PVD technique.
Coating Process
Diamond and TC burs were placed in an ultrasonic bath containing distilled water for cleaning and were then rinsed with acetone and ethyl alcohol ( > 95%). Next, they were mounted on respective fixtures on the carousel of the vacuum coating machine (Sevin Plasma, Iran) such that each bur rotated around its own central axis and the central axis of the carousel at a speed of 50 rpm and 4 rpm, respectively, and had averagely 30 cm distance from the sputtering source. The chamber was then closed and vacuumed to 3 × 10-5 mbar pressure. Vacuuming was performed in three steps by the rotary mechanical (to 5 × 10-1 mbar pressure), roots vacuum (to 5 × 10-3 mbar pressure), and diffusion (to 5 × 10-5 mbar ) pumps. Then, the burs were heated to 350°C and ionized by the input of argon gas into the vacuum chamber at 20 000 V, and the pressure was adjusted at 10-2 mbar. Next, the burs were bumbarded with argon gas ions to eliminate the impurities from the surface. To apply the DLC coating, first, the gas input was discontinued to adjust the pressure at 3 × 10-5 mbar. Subsequently, acetylene gas in the 8:2 ratio was used to adjust the pressure at 2 × 10-2 mbar. The sputtering source of the device on which a chromium target was installed, started chromium sputtering by applying 1 kV voltage, and the Cr-DLC phase was applied on each bur. By increasing the acetylene input, the rate of Cr sputtering decreased, while the deposition of the DLC phase represented an increase. This process was continued for 120 minutes to reach the required coating thickness of 3.5 µm.
Pin-on-Disc Wear Test
First, each bur was weighed by a digital scale with 0.001 mg accuracy. Next, the burs were placed in the respective fixtures of the pin-on-disc device (Razi Institute, Iran) operating at a crosshead speed of 1.5 m/s with a 15 mm radius, 20 m distance, 954 rpm, and 500 N applied load. The distance, speed, and load were recorded in a computer, and the burs underwent wear. After the completion of the wear test, the burs were weighed again, and the difference between their primary and secondary weight was calculated and reported in milligrams (mg).
Qualitative Assessment
The burs were also inspected and photographed under a stereomicroscope (Iran) at × 4 and × 10 magnifications at baseline (before testing), halfway through the wear test, and after the completion of the test.
Statistical Analysis
Data were analyzed by SPSS, version 21. The normal distribution of data was evaluated by the Shapiro-Wilk test. Considering the normal distribution of the data (P > 0.05), the two-way analysis of variance (ANOVA) and Tukey’s test were used for general and pairwise comparisons of the mean wear of the study groups, respectively. The level of significance was set at 0.05.
Results
Quantitative Findings
Table 1 presents the mean and standard deviation of the weight loss of burs after the wear test compared with the baseline. Two-way ANOVA results showed the significant effect of coating on wear (P < 0.05); however, the effect of type of bur and their interaction effect were not significant on wear (P > 0.05, Table 2). The pairwise comparisons of the groups regarding wear by Tukey’s test demonstrated significantly lower wear of coated burs than non-coated burs (P < 0.05, Table 3).
Table 1.
Mean ± SD of Weight Loss of Burs After the Wear Test Compared With Baseline in Milligrams (n = 15)
Group
|
Mean±SD of Wear
|
Carbide burs without coating |
1.52 ± 6.24 |
Carbide burs with coating |
0.31 ± 0.4 |
Diamond burs without coating |
2.3 ± 6.7 |
Diamond burs with coating |
0.71 ± 0.98 |
Note. SD: standard deviation.
Table 2.
Effect of Type of Bur and Presence of Coating on Wear of Burs Using Two-Way ANOVA
Source
|
Sum of Squares
|
Degree of Freedom
|
Mean Square
|
F
|
P Value
|
Coefficient |
774.00 |
1 |
774.00 |
370.66 |
P < 0.05 |
Coating |
505.18 |
1 |
505.18 |
241.92 |
P < 0.05 |
Bur type |
4.42 |
1 |
4.42 |
2.12 |
0.151 |
Coating × Bur type |
0.02 |
1 |
0.02 |
0.01 |
0.922 |
Error |
116.93 |
56 |
2.08 |
|
|
Total |
1400 |
60 |
|
|
|
Note. ANOVA: Analysis of variance.
Table 3.
Pairwise Comparisons of the Wear (Weight Loss) of Burs in the Four Groups by the Tukey’s Test
|
Compared Groups
|
Mean Difference
|
P Value
|
Tukey’s test |
Diamond burs without coating |
TC burs without coating |
0.50 |
0.772 |
Diamond burs with coating |
5.76 |
< 0.05 |
TC burs with coating |
6.34 |
< 0.05 |
TC burs without coating |
Diamond burs with coating |
5.26 |
< 0.05 |
TC burs with coating |
5.84 |
< 0.05 |
Diamond burs with coating |
TC burs with coating |
0.58 |
0.691 |
Note. TC: Tungsten carbide.
Qualitative Findings
Figure 1 shows a sound diamond particle on the bur surface prior to wear at × 4 and × 10 magnifications. Figure 2 illustrates the same site of the bur surface after the wear test. As shown, all diamond particles have been detached from the surface, confirming the effects of wear on the metal surface.
Figure 1.
Stereomicroscopic Image of the Surface of a Diamond Bur at × 10 Magnification
Figure 1.
Stereomicroscopic Image of the Surface of a Diamond Bur at × 10 Magnification
Figure 2.
(A) Surface of a Diamond Bur Prior to the Wear Test at × 4 Magnification and (B) Surface of a Diamond Bur Prior to the Wear Test at × 10 Magnification
Figure 2.
(A) Surface of a Diamond Bur Prior to the Wear Test at × 4 Magnification and (B) Surface of a Diamond Bur Prior to the Wear Test at × 10 Magnification
Figure 3 depicts the stereomicroscopic image of the cutting tip and blades of a TC bur prior to wear at × 10 magnification. Figure 4 displays the surface of the TC bur after the completion of the test.
Figure 3.
(A) Surface of a Diamond Bur After the Wear Test at × 4 Magnification and (B) Surface of a Diamond Bur After the Wear Test at × 10 Magnification
Figure 3.
(A) Surface of a Diamond Bur After the Wear Test at × 4 Magnification and (B) Surface of a Diamond Bur After the Wear Test at × 10 Magnification
Figure 4.
Stereomicroscopic Image of the Cutting Tip and Blades of a Coated Carbide Bur Prior to Wear at × 10 Magnification
Figure 4.
Stereomicroscopic Image of the Cutting Tip and Blades of a Coated Carbide Bur Prior to Wear at × 10 Magnification
Discussion
This study assessed the effect of DLC coating applied by the PVD technique on the wear of diamond and TC dental burs. DLC coating is more effective than other coatings such as alumina because diamond is harder than alumina and has excellent physical and chemical properties (25). Thus, DLC coating was applied on burs by the PVD technique in the present study as described by Stoessel and Stoessel (16) to provide a hard coating, which is resistant to wear and corrosion. The thickness of the applied coating was selected to be 3.5 µm in the current study, which was similar to a study by Tuffy et al (26). They concluded that titanium nitride coating with 3.5 µm thickness had the best performance, and other thicknesses had an inferior performance due to their higher stress level, which resulted in the fracture of the coatings due to the crushing of the carbide cutting blade. However, such problems did not exist in the application of 3.5 µm coating. On the other hand, Dorner et al (27) found that DLC coatings with 0.7-3.5 µm thickness had no significant effect on the wear properties of biomedical Ti6Al4V alloy because microstructural differences resulted in a heterogenous DLC coating. Their results differ from the present findings probably due to the use of an abrasive device that used water pressure and different elements on the alloy surface. Qin et al (13) evaluated the effect of the thickness of diamond coating applied on the cutting instruments and reported that despite higher stresses at the interface, the instruments with thicker coatings had higher durability and longer clinical service.
The results of the present study revealed that the application of DLC coating generally improved the quality and durability of burs compared with non-coated burs, which is in agreement with the findings of Ahmed et al (5) and Qin et al (13). They found that the coating of burs prevented the reduction in the quality of the cutting blades and increased their durability. The present results are also in line with those of Borges et al (28), implying that the steriliziation process and clinical application of diamond burs resulted in a reduction in diamond particles, and subsequent reductions in the quality and clinical service of burs and hygienic contamination. They added that the coating of burs significantly decreased the detachment of diamond particles from the burs.
The current results revealed significantly lower wear of coated TC and diamond burs compared with non-coated control burs, indicating that coating significantly increased the corrosion resistance and wear resistance of burs. This result conforms to the findings of Sein et al (29), demonstrating that coating significantly increased the cutting efficiency of burs. The application of the coating on the bur surface decreased the wear of burs by 3 times, compared with non-coated burs, prevented their early wear, and improved their clinical service and function.
The present results also represented that the mean wear of diamond burs was higher than that of TC burs, which is because the main cause of wear of diamond burs is the detachment of diamond particles from the surface. However, TC burs do not have such particles, and the entire coating should be removed for the TC bur to undergo wear. This finding is in accordance with the results of Tang et al (30), highlighting that stress can cause microcracks and fracture in burs, while the coating of burs results in the generation of compressive stresses.
Stereomicroscopic assessments in the present study revealed that the diamond particles were loosely attached to the surface of diamond burs in some areas due to incorrect storage or transfer conditions, which can decrease the longevity of burs. This finding is in line with the results of Fuentes et al (31). They concluded that in non-coated burs, diamond particles are detached from the surface and decrease the longevity of burs, leading to wear of the empty space left on the bur surface and the formation of a smear layer of bur in the form of a small spot on the tooth surface. However, coating improves the resistance of burs.
The results of the current study showed that the wear of coated TC burs in the second half of the test was extremely lower than that in the first half, which is in complete agreement with the quantitative test results. This finding is due to a continuous increase in the cross-sectional area of the bur due to friction. As the cross-sectional area increases, the stress applied to the surface decreases because the surface area has an inverse correlation with the applied stress. This result corroborates the findings of Sein et al (32); they assessed the efficacy of the coating of TC burs and found that stress distribution at the tip of the bur was higher than that in other areas and caused a higher temperature rise at the tip compared with other areas. Harano et al (33) used microscopic images to assess the effect of wear on different coatings of dental burs. They reported that burs with CVD diamond coating had the highest longevity and clinical service, which is in conformity with the results of qualitative assessments in the present study.
The qualitative assessment of burs in the current study revealed the detachment of diamond particles and the impression of the path of wear on the surface of diamond burs, as well as the irregularities and impression of the wear path on the surface of carbide burs. Pournasiri et al (34) evaluated the adverse effects of sterilization cycles on diamond and TC burs. Microscopic assessments demonstrated a reduction in the number of diamond particles in diamond burs after each sterilization cycle. They also found irregularities on the surface of TC burs after repeated sterilization cycles, which is consistent with the qualitative findings of the present study. In diamond burs, friction results in the detachment of diamond particles from the surface and corrosion of the underlying metal (31). Qualitative images enable the qualitative assessment of the degradation level of diamond particles and underlying metal at each step of testing; otherwise, it cannot be found that weight loss occurs due to wear of which part of the bur.
The qualitative assessment of the surface of burs in addition to the quantitative assessment was a strength of this study. Quantitative assessment only indicates weight loss, while qualitative assessment reveals the parts that experience maximum degradation due to wear.
On the other hand, this study had some limitations. Despite its in vitro design, which limits the generalization of the results to the clinical setting, this study only assessed the effect of coating on the wear of burs. Future studies are required to investigate the effect of coating on fracture resistance, hardness, and cutting efficiency of burs. In addition, the pin-on-disc wear test was used in the present study. Future studies are needed to evaluate the adverse effects of wear of burs on enamel and dentin. Furthermore, clinical studies are required to assess the impact of the duration of the use of burs on their surface quality and optimal clinical service of coated and non-coated burs.
Conclusions
The DLC coating of the diamond and TC dental burs by the PVD technique increased their wear resistance regardless of the bur type.
Authors’ Contribution
Conceptualization: Bahareh Asgartooran.
Data curation: Zahra Khamverdi.
Formal analysis: Zahra Khamverdi.
Investigation: Faezeh Soury.
Methodology: Shiva Kavousinejad, Faezeh Soury.
Project administration: Bahareh Asgartooran.
Resources: Shiva Kavousinejad.
Supervision: Zahra Khamverdi, Bahareh Asgartooran.
Validation: Zahra Khamverdi.
Visualization: Bahareh Asgartooran.
Writing – original draft: Shiva Kavousinejad.
Writing – review & editing: Shiva Kavousinejad.
Competing Interests
The authors declare that they have no conflict of interests.
Ethical Approval
The study was approved by the Ethics Committee of Hamadan University of Medical Sciences (Code: IR.UMSHA.REC.1399.237).
References
- Gwoździk M, Bałaga Z, Wróbel D, Nitkiewicz Z. Evaluation of wear degree of rotational instruments with diamond coat. Compos Theory Pract 2017; 17(4):216-20. [ Google Scholar]
- Ahmed W, Sein H, Ali N, Gracio J, Woodwards R. Diamond films grown on cemented WC–Co dental burs using an improved CVD method. Diam Relat Mater 2003; 12(8):1300-6. doi: 10.1016/s0925-9635(03)00074-8 [Crossref] [ Google Scholar]
- Pines MS, Schulman A. Characterization of wear of tungsten carbide burs. J Am Dent Assoc 1979; 99(5):831-3. doi: 10.14219/jada.archive.1979.0403 [Crossref] [ Google Scholar]
- Norman RD, Mehra RV, Swartz ML, Phillips RW. Effects of restorative materials on plaque composition. J Dent Res 1972; 51(6):1596-601. doi: 10.1177/00220345720510061601 [Crossref] [ Google Scholar]
- Ahmed W, Sein H, Jackson M, Rego C, Hassan IU, Subramani K. Surface engineering of dental tools with diamond for enhanced life and performance. In: Subramani K, Ahmed W, eds. Emerging Nanotechnologies in Dentistry. 2nd ed. William Andrew Publishing; 2018. p. 251-88. 10.1016/b978-0-12-812291-4.00013-3.
- Ahmed W, Sein H, Jackson M, Polini R. Chemical vapour deposition of diamond films onto tungsten carbide dental burs. Tribol Int 2004; 37(11-12):957-64. doi: 10.1016/j.triboint.2004.07.013 [Crossref] [ Google Scholar]
- Strzelczak K. Quality control of a metallic dental bur with a diamond coating. Prod Eng Arch 2018; 19(19):26-9. doi: 10.30657/pea.2018.19.06 [Crossref] [ Google Scholar]
- Klimek L, Kochanowski M, Romanowicz M. Abrasive wear of diamond-coated dental burs and its impact on the parameters of the finished surface. J Superhard Mater 2007; 29(3):181-4. doi: 10.3103/s1063457607030148 [Crossref] [ Google Scholar]
- Lin M, Xu F, Lu TJ, Bai BF. A review of heat transfer in human tooth--experimental characterization and mathematical modeling. Dent Mater 2010; 26(6):501-13. doi: 10.1016/j.dental.2010.02.009 [Crossref] [ Google Scholar]
- Soori A, Kowsary F, Kasraei S. Experimental estimation of the emissivity of human enamel and dentin. Infrared Phys Technol 2020; 106:103234. doi: 10.1016/j.infrared.2020.103234 [Crossref] [ Google Scholar]
- Zach L, Cohen G. Pulp response to externally applied heat. Oral Surg Oral Med Oral Pathol 1965; 19:515-30. doi: 10.1016/0030-4220(65)90015-0 [Crossref] [ Google Scholar]
- Jalali F, Ghatee M, Hashemian SM. Effect of TiN/TiCN/Al2O3 multilayer coating on tungsten carbide drill on drilling process of spherodized cast iron. Modares Mechanical Engineering 2016;15(12):443-9. [Persian].
- Qin F, Chou YK, Nolen D, Thompson RG. Coating thickness effects on diamond coated cutting tools. Surf Coat Technol 2009; 204(6-7):1056-60. doi: 10.1016/j.surfcoat.2009.06.011 [Crossref] [ Google Scholar]
- Khamverdi Z, Yarmohammadi E, Tolaminejad B, Barghilashgari A. Effect of sterilization (autoclave) on wear of tungsten-carbide burs coated with diamond particles with different thicknesses. Avicenna J Dent Res 2021; 13(4):119-23. doi: 10.34172/ajdr.2021.23 [Crossref] [ Google Scholar]
- He Y, Apachitei I, Zhou J, Walstock T, Duszczyk J. Effect of prior plasma nitriding applied to a hot-work tool steel on the scratch-resistant properties of PACVD TiBN and TiCN coatings. Surf Coat Technol 2006; 201(6):2534-9. doi: 10.1016/j.surfcoat.2006.04.064 [Crossref] [ Google Scholar]
- Bunshah RF, Weissmantel C. Handbook of hard coatings: Noyes Publications Park Ridge, NJ; 2001.
- Bunshah RF, Weissmantel C. Handbook of Hard Coatings. Park Ridge, NJ: Noyes Publications; 1999.
- Bhatt A, Attia H, Vargas R, Thomson V. Wear mechanisms of WC coated and uncoated tools in finish turning of Inconel 718. Tribol Int 2010; 43(5-6):1113-21. doi: 10.1016/j.triboint.2009.12.053 [Crossref] [ Google Scholar]
- Chang CL, Wang DY. Microstructure and adhesion characteristics of diamond-like carbon films deposited on steel substrates. Diam Relat Mater 2001; 10(8):1528-34. doi: 10.1016/s0925-9635(01)00382-x [Crossref] [ Google Scholar]
- Corazza PH, de Moura Silva A, Cavalcanti Queiroz JR, Salazar Marocho SM, Bottino MA, Massi M. Carbon film coating of abutment surfaces: effect on the abutment screw removal torque. Implant Dent 2014; 23(4):434-8. doi: 10.1097/id.0000000000000111 [Crossref] [ Google Scholar]
- Diez JS, Brigagão VC, Cunha L, Neves AC, da Silva-Concilio LR. Influence of diamondlike carbon-coated screws on the implant-abutment interface. Int J Oral Maxillofac Implants 2012; 27(5):1055-60. [ Google Scholar]
- Liao YS, Lin TC, Lai CY, Chen YL, Chang HH, Lin CP. Cutting performance of diamond-like carbon coated tips in ultrasonic osteotomy. J Dent Sci 2014; 9(1):63-8. doi: 10.1016/j.jds.2013.02.014 [Crossref] [ Google Scholar]
- Kobayashi S, Ohgoe Y, Ozeki K, Sato K, Sumiya T, Hirakuri KK. Diamond-like carbon coatings on orthodontic archwires. Diam Relat Mater 2005; 14(3-7):1094-7. doi: 10.1016/j.diamond.2004.11.036 [Crossref] [ Google Scholar]
- Sein H, Ahmed W, Jackson M, Woodwards R, Polini R. Performance and characterisation of CVD diamond coated, sintered diamond and WC–Co cutting tools for dental and micromachining applications. Thin Solid Films 2004; 447-448:455-61. doi: 10.1016/j.tsf.2003.08.044 [Crossref] [ Google Scholar]
- Ahmed W, Rego CA, Cherry R, Afzal A, Ali N, Hassan IU. CVD diamond: controlling structure and morphology. Vacuum 2000; 56(3):153-8. doi: 10.1016/s0042-207x(99)00147-5 [Crossref] [ Google Scholar]
- Tuffy K, Byrne G, Dowling D. Determination of the optimum TiN coating thickness on WC inserts for machining carbon steels. J Mater Process Technol 2004; 155-156:1861-6. doi: 10.1016/j.jmatprotec.2004.04.277 [Crossref] [ Google Scholar]
- Dorner A, Schürer C, Reisel G, Irmer G, Seidel O, Müller E. Diamond-like carbon-coated Ti6Al4V: influence of the coating thickness on the structure and the abrasive wear resistance. Wear 2001; 249(5-6):489-97. doi: 10.1016/s0043-1648(01)00587-7 [Crossref] [ Google Scholar]
- Borges CF, Magne P, Pfender E, Heberlein J. Dental diamond burs made with a new technology. J Prosthet Dent 1999; 82(1):73-9. doi: 10.1016/s0022-3913(99)70130-7 [Crossref] [ Google Scholar]
- Sein H, Ahmed W, Rego CA, Jones AN, Amar M, Jackson M. Chemical vapour deposition diamond coating on tungsten carbide dental cutting tools. J Phys Condens Matter 2003; 15(39):S2961-7. doi: 10.1088/0953-8984/15/39/019 [Crossref] [ Google Scholar]
- Tang W, Wang Q, Wang S, Lu F. Adherent diamond coatings on cemented carbide substrates with different cobalt contents. Diam Relat Mater 2001; 10(9-10):1700-4. doi: 10.1016/s0925-9635(01)00398-3 [Crossref] [ Google Scholar]
- Fuentes GG, Almandoz E, Rodríguez RJ, Dong H, Qin Y, Mato S. Vapour deposition technologies for the fabrication of hot-forming tools: a review. Manuf Rev 2014; 1:20. doi: 10.1051/mfreview/2014019 [Crossref] [ Google Scholar]
- Sein H, Ahmed W, Rego C. Application of diamond coatings onto small dental tools. Diam Relat Mater 2002; 11(3-6):731-5. doi: 10.1016/s0925-9635(01)00579-9 [Crossref] [ Google Scholar]
- Harano K, Satoh T, Sumiya H. Cutting performance of nano-polycrystalline diamond. Diam Relat Mater 2012; 24:78-82. doi: 10.1016/j.diamond.2011.11.005 [Crossref] [ Google Scholar]
- Pournasiri I, Neshandar Asli H, Abbaspour R, Maleki D. The effects of different sterilization methods on the uniformity of diamond dental burs and changes in the cutting edge carbide burs. Journal of Dentomaxillofacial Radiology, Pathology and Surgery 2018; 7(2):77-84. doi: 10.32598/3dj.7.2.77 [Crossref] [ Google Scholar]